Chapter 2 Compatible Reprocessing Methods and Chemical Agents
13
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
• Allow the packages to dry within the autoclave, using the
autoclave’s drying cycle (if available) or by opening the door
of the autoclave and allowing the packages to air-dry.
Handling a wet package can compromise its sterility.
Process Parameters
Prevacuum Temperature 132 – 134C
(270 – 274F)
Exposure time 5minutes
Table 2.3 Steam sterilization (autoclaving) exposure parameters