Chapter 4 Operation
MAINTENANCE UNIT MU-1
4.2 Remove water from the Air/Water Channel (For
endoscope with Air/Water channel only)
Do not remove the AW Channel Cleaning Adapter and the
Alcohol Container’s nozzle until the Power Switch is OFF
”. If the AW Channel Cleaning Adapter or Alcohol
Container’s nozzle are removed from the endoscope while
the Power Switch is ON “ ”, water or 70% ethyl or isopropyl
alcohol will splatter.
Flushing the channels with 70% ethyl or isopropyl alcohol
facilitates drying inside the Air/Water channel.
When Irrigating With 70% Ethyl or Isopropyl Alcohol
1.
Turn the Power Switch ON “ ”.
2. Depress the button on the AW Channel Cleaning Adapter for approximately
10 seconds to irrigate the channel with 70% ethyl or isopropyl alcohol (see
Figure 4.1).
3. Release the button and perform insufflation for approximately 10 seconds
until no more 70% ethyl or isopropyl alcohol exits the endoscope.
4. Turn the Power Switch OFF “ ”.
5. Remove the Alcohol Container’s nozzle from the Air/Water Port on the Light
Guide Connector.
6. Perform drying the Air/Water channel according to the next page.
Figure 4.1