ISSUE6 6-3 Safety Check
MU-1
4. Safety Check Sheet
Electrical Safety Inspection
Date of inspection
Product MU-1 Inspected by
Serial number Checked by
Item Criteria Result
2-1-1 External leak current
Normal Condition
100 A or less
Single Fault
500 A or less
2-1-2 Grounding leak current
Normal Condition
500 A or less
Single Fault
1000 A or less
2-1-3 Grounding Resistance Between the external case and grounding line. 0.2 Ω or less