9 Compatible equipment
Olympus recommends the use of equipment as listed in this
chapter. If combinations are used that are not listed in this
chapter, the user takes the full responsibility.
Future equipment may also be compatible. For more information,
contact an Olympus representative.
Some of the products listed in this chapter may not be available
in all sales territories.
CAUTION
Risk of injury to the patient and/or the user
When using endoscopes and light-guide cables with an
inappropriate combination of the maximum insertion portion width
and ber bundle diameter, the connectors may become hot.
• Only use endoscopes and light-guide cables with the
appropriate combination of the maximum insertion portion
width and ber bundle diameter.
• For the appropriate maximum insertion portion width, refer to
the chapter “Technical data” on page 45.
• For the appropriate ber bundle diameter, refer to the
respective instructions for use of the light-guide cable.
• Use the instructions in this document if the instructions differ
from the instructions for use of the light-guide cable.
For reasons of safety, the active diameter of the light-guide cable
used in combination with the endoscope must be as specied:
Diameter of the endoscope ≤4.1 mm: active diameter of the
light-guide cable ≤3 mm.
Using endoscopes in combination with light-guide cables of other
manufacturers may lead to an increase in temperature of the
connectors.