EasyManua.ls Logo

Omega 3200 - D) Connections for the 9 Pin D Sub Connector

Omega 3200
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
M-4271/0707, pg. 11 of 26
d) Connections For The 9 Pin D Sub Connector
Using a suitable mating connector the pins of the integrated connector
should be wired as follows:
Connecting To The Integrated 9 Pin Connector
PIN 3 should be connected to the Positive of the power source.
PIN 4 should be connected to the Negative ( Ground ) of the power
source.
PIN 2 provides the signal output and should be connected to the positive
terminal of the display, data acquisition system or voltmeter.
PIN 8 is the signal negative (ground) and should be connected to the
negative (Ground) terminal of the display, data acquisition system or
voltmeter.
Pin 6 is the input signal and should be connected to the positive terminal
of the voltage source. The (0-5VDC) voltage control signal should be
supplied from a low impedance source.
Pin 7 is the input signal negative (ground) and should be connected to
the negative (Ground) terminal of the voltage source.
Pins 1, 5, and 9 are not used.
Caution: Avoid high voltage static discharges to the input signal
connection. Do not short the input/output signal pins or allow
them to contact the power connections at any time. DAMAGE
WILL RESULT!