2.4  Installing and Changing Chart Paper
Make sure the Sensor probe is plugged into the Recorder when power is on. The 
Factory set up is for 7 Day Chart Speed & °F Chart Scale for CTXL-TRH.
To install or change the same Chart paper:
• Open the Recorder door & remove the magnetic hub that holds the paper.
• Place the new Chart paper.
• Rotate the paper, until the present time lines up with the Time reference arrow 
as shown in Fig 2.8.
• Place the magnetic hub over the paper & close the recorder door. This can be 
done with or without power turned on.
2.4.1 Changing Chart Speed & Scale
You can change the Chart speed & scale as follows:
• Make sure the unit is turned on by pressing the Power Key.
• Open the Recorder door. The pens will go to the Home position (Bottom of 
scale) and the display will show “door oPEn” as well as the existing Chart 
Speed (7 Days) and Chart scale (°F).
• Press the Chart Speed key
 
SPEED
SET
LIGHT
NO
POWER
CONFIG YES MODE
 
to change to 1 or 32 day chart.
• Press the Chart Scale key
SCALE
CLOCK SPEED SET
LIGHT
NO
POWER
CONFIG YES MODE
 to change to °C if needed (CTXL-TRH only).
• Install the new Chart paper as explained in the previous paragraph
• Close the Recorder door. The display will show “dAtA rSt?”
• Press 
°C
°F
SCALE
CLOCK SPEED SET
LIGHT
NO
POWER
CONFIG
YES
 key to reset (erase) previous stored data in memory. The unit 
 beeps for verification.
• Press 
°C
°F
SCALE
CLOCK SPEED SET
LIGHT
NO
 key to continue recording without resetting (erasing) previous 
         stored data in memory.
• If no keys are pressed, the unit will resume recording after 5 seconds without 
erasing previous stored data in memory
°F
7D
°C
1D
Open
Recorder
Door
Close
Recorder
Door
Close
Recorder
Door
Change
Chart
Speed &
Scale
Delete
Old Data
Real
Time Mode
Press
Press
Unit Beeps
or no key
Figure 2-7 Changing Chart Speed & Scale Flow Chart
NO
YES
Figure 2-7.  Changing Chart Speed and Scale
Flow Diagram
Setting Up the Recorder
2