PC Interface Software
3
3-2
3.2  Re-scaling One or Both Pens on Chart Paper
You can re-scale one or both pens to any range within the operating input range. 
Here are the commands:
NOTE
The rescaling of the Temperature Pen is based on Degree F 
scale on both CTXL-TRH and CTXL-DTC Models.
 Command  Description    Example
 PUT  Change Chan 1 Pen (Temp)   PUT90; Pen Scale set 
  top chart value 
 PUB  Change Chan 1 Pen (Temp)   PUB50; Pen Scale set 
  bottom chart value
 PLT  Change Chan 2 Pen (Humidity)  PLT80; Pen Scale set 
  top chart value 
 PLB  Change Chan 2 Pen (Humidity)   PLB40; Pen Scale set   
   bottom chart value 
 PR0  Issue the new Pen scales values   PR0; Rescaled 
  to the recorder
 PS  Get the Pen scales for   PS;0090,0050,0080,0040 
  Chan 1 & 2
 PR1  Re-scale the Pens to the   PR1; Rescaled 
  standard Factory scales 
 In the above example, we are re-scaling both Pens as follows:
Temperature Pen: 50 to 90ºF (Standard: 0 to 120ºF)
Humidity Pen: 40 to 80% RH (Standard: 0 to 100%RH)
After issuing the PR0 command, the new Pen scale values will take affect, otherwise 
the previous scale values will remain. The Chart icon comes on to indicate new chart 
scale. The new scale is stored in the non-volatile memory. To restore the Pen scales 
back to the Factory settings, you need to issue the PR1 command.
The rescaling value can be a negative number on model CTXL-DTC. For example, 
you can issue a PC command: PUB-100 meaning the bottom of chart for channel 1 
starts at -100.