PLATINUM
TM
Series Meters Quick Start Guide
5
Wiring Instructions
Caution: Do not connect power to your device until you have completed all input
and output connections. Failure to do so may result in injury!
Use Copper Conductors only
(rated for 75
C min.)
for Power Connections
3.2 Connecting Power
Connect the main power connections to pins 7 and 8 of the 8-pin power / output connector as shown in
Figure 3.
Figure 3 Main Power Connections
For the low-voltage power option, maintain the same degree of
protection as the standard high-
voltage input power units (90–240 Vac) by
using a Safety Agency Approved DC or AC source with
the same Overvoltage Category and pollution degree as the standard AC unit (90–300 Vac).
The Safety European Standard 2014/35/EU for measurement, control, and laboratory equipment
requires that fuses must be specified based on IEC127. This standard specifies the letter code “T”
for a Time-lag fuse.
110 ~ 300 Vdc
110 - 300 Vdc