36
appears or a
measurement
cannot be complete
after the arm cuff
inflates.
You move or talk during
a measurement and the
arm cuff does not inflate
sufficiently.
Remain still and do not
talk during a
measurement. If “E2”
appears repeatedly,
inflate the arm cuff
manually until the
systolic pressure is 30 to
40 mmHg above your
previous readings. Refer
to sub-section 3.1.
The systolic pressure is
above 210 mmHg and a
measurement cannot be
taken.
Inflate the arm cuff
manually until the
systolic pressure is 30 to
40 mmHg above your
previous readings. Refer
to sub-section 3.1.
appears
The arm cuff is inflated
exceeding the maximum
allowable pressure.
Do not touch the arm cuff
and/or bend the air tube
while taking a
measurement.
If inflating the arm cuff
manually, refer to the
end of sub-section 3.1.
appears
You move or talk during
a measurement.
Vibrations disrupt a
measurement.
Remain still and do not
talk during a
measurement.
Display/Problem Possible Cause Solution

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron 5 Series and is the answer not in the manual?

Omron 5 Series Specifications

General IconGeneral
ModelBP7200
TypeUpper Arm
Memory Storage60 readings
DisplayDigital LCD
Irregular Heartbeat DetectionYes
Measurement MethodOscillometric
Measurement Range (Pressure)0 to 299 mmHg
Body Movement DetectionYes
Power Source4 AA batteries
Measurement Range (Pulse Rate)40 to 180 beats/min.
Average ReadingYes
Fits Arm Circumference22-42 cm
Users2 users