10. Specifications
41
•These specifications are subject to change without notice.
•This monitor complies with the requirements of ISO 81060-2:2013.
In the clinical validation study, K5 was used on 85 subjects for
determination of diastolic blood pressure.
•This monitor has not been validated for use on pregnant patients.
•IP classification is degrees of protection provided by enclosures in
accordance with IEC 60529. This monitor and optional AC adapter
are protected against solid foreign objects of 0.5 inch (12.5 mm)
diameter and greater such as a finger. The optional AC adapter is
protected against vertically falling water drops which may cause
issues during a normal operation.
Weight
Monitor: Approximately 16.2 oz. (460 g) not
including batteries
Arm cuff: Approximately 5.8 oz. (163 g)
Dimensions
Monitor: Approximately 7.5" (w) × 3.1" (h) ×
4.1" (l) (191 mm × 79 mm × 105 mm)
Arm cuff: Approximately 5.7" × 20.9" (air
tube: 29.5") (145 mm × 532 mm (air tube:
750 mm))
Cuff circumference
applicable to the
monitor
9" to 17" (220 to 420 mm)
Memory Stores up to 60 readings per user
Contents
Monitor, Easy Wrap ComFit Cuff, 4 “AA”
batteries, instruction manual, quick start
guide
Protection against
electric shock
Internally powered ME equipment (When
using only batteries)
Class II ME equipment (Optional AC
adapter)
Applied part Type BF (arm cuff)