24
3. Using Your Monitor without a Smart Device
If your systolic pressure is more than 210 mmHg
If your systolic pressure is more than 210 mmHg, take a
measurement manually.
After the arm cuff starts to inflate, press and hold the [START/STOP]
button until the monitor inflates 30 to 40 mmHg higher than your
expected systolic pressure.
Note
Do not inflate above 299 mmHg.
Caution
Inflating to a higher pressure than necessary may result in bruising where the
arm cuff is applied.