Introduction
Thank you for purchasing the OMRON Upper Arm Blood Pressure Monitor.
Your new upper arm blood pressure monitor uses the oscillometric method of 
blood pressure measurement. This means the monitor detects your blood 
movement through your brachial artery and converts the movements into a 
digital reading. An oscillometric monitor does not need a stethoscope so the 
monitor is simple to use.
Intended Use
This device is a digital monitor intended for use in measuring blood pressure and 
pulse rate in adult patient population. The device detects the appearance of 
irregular heartbeats during measurement and gives a warning signal with 
readings. It is mainly designed for general household use. 
Important Safety Information
Please read the Important Safety Information in this instruction manual 
before using the unit.
Please follow this instruction manual thoroughly for your safety.
Please keep for future reference. For specific information about your own blood 
pressure, CONSULT YOUR DOCTOR.
Warning
Indicates a potentially hazardous situation which, if not 
avoided, could result in death or serious injury.
• DO NOT adjust medication based on measurement results from this blood 
pressure monitor. Take medication as prescribed by your doctor. Only a doctor 
is qualified to diagnose and treat high blood pressure.
• DO NOT use this monitor on an injured arm or an arm under medical 
treatment.
• DO NOT wear the arm cuff on your arm while on an intravenous drip or blood 
transfusion.
• DO NOT apply or inflate the arm cuff around any body parts other than your 
upper arm.
• DO NOT use this monitor in areas containing high frequency (HF) surgical 
equipment, magnetic resonance imaging (MRI) equipment, computerized 
tomography (CT) scanners or in oxygen rich environments.
• DO NOT use this monitor on infants, toddlers, children or persons who cannot 
express themselves.
• Consult your doctor before using this monitor if you have common arrhythmias 
such as atrial or ventricular premature beats or atrial fibrillation; arterial 
sclerosis; poor perfusion; diabetes; pregnancy; pre-eclampsia; or renal 
disease.
• Always consult your doctor. Self-diagnosis of measurement results and 
self-treatment are dangerous.
• This product contains small parts that may cause a choking hazard if 
swallowed by children, toddlers or infants.
• Keep the BATTERY out of the reach of children, toddlers and infants.
• This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use 
this product in locations where RF is restricted, such as on an aircraft or in 
hospitals. 
Caution
Indicates a potentially hazardous situation which, if not 
avoided, may result in minor or moderate injury to the 
user or patient or damage to the equipment or other 
property.
• Stop using this monitor and consult your doctor if you experience skin irritation 
or discomfort.
• Consult your doctor before using this monitor on an arm with an arterio-venous 
(A-V) shunt.
• Note that PATIENT motion, trembling, shivering may affect the measurement 
reading.
• Consult your doctor before using this monitor if you have had a mastectomy.
• Consult your doctor before using the monitor if you have severe blood flow 
problems or blood disorders as cuff inflation can cause bruising.
• Do not take measurements more often than necessary because bruising, due 
to blood flow interference, may occur.
• Remove the arm cuff if it does not start deflating during the measurement.
• Do not use this monitor for any purpose other than measuring blood pressure.
• Do not use a mobile phone or other electrical devices that emit electromagnetic 
fields near this monitor except when transferring your readings to your smart 
device. This may result in incorrect operation of this monitor.
• Do not disassemble or attempt to repair the monitor or components. This may 
cause an inaccurate reading.
• Do not use in a location where there is moisture or a risk of water splashing the 
monitor. This may damage the monitor.
• Do not use this monitor in a moving vehicle such as in a car or on an aircraft. 
Do not drop, or subject this monitor to strong shocks or vibrations.
• Do not use this monitor in places with high or low humidity or high or low 
temperatures. See the “6. Specifications” section of the manual.
• Please check, by observation of the arm being used, that the monitor is not 
causing any impairment of patient blood circulation.
• Do not use this monitor in high-use environments such as medical clinics or 
doctor offices.
• Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for 
30 minutes before taking a measurement.
• Rest for at least 5 minutes before taking the measurement.
• Remove tight-fitting or thick clothing from your arm while taking a 
measurement.
• Remain still and do not talk while taking a measurement.
• Read and follow the “Important information regarding Electro Magnetic 
Compatibility (EMC)” in the EMC information provided with this device.
• Read and follow the “Correct Disposal of This Product” in “6. Specifications” 
when disposing of the device and any used accessories or optional parts.
Data Transmission
• Do not replace the battery while your reading is being transferred to your smart 
device. This may result in the incorrect operation of your monitor and failure to 
transfer your blood pressure data.
• Do not place integrated circuit cards, magnets, metal objects, or other devices 
that emit electromagnetic fields near this monitor while your reading is being 
transferred to your smart device. This may result in the incorrect operation of 
your monitor and failure to transfer your blood pressure data.
Battery Handling and Usage
• Do not insert the batteries with their polarities incorrectly aligned.
• Use only 4 “AAA” alkaline batteries with this monitor. Do not use other types of 
batteries. Do not use new and used batteries together.
• Remove the batteries if the monitor will not be used for a long period of time.
• If battery fluid should get in your eyes, immediately rinse with plenty of clean 
water. Consult a doctor immediately.
• If battery fluid should get on your skin, wash your skin immediately with plenty 
of clean, lukewarm water. If irritation, injury or pain persists, consult a doctor.
• Do not use batteries after their expiration date.
• Periodically check your batteries to ensure they are in good working condition.
• Only use batteries specified for this monitor. Use of unsupported batteries may 
damage and/or may be hazardous to the monitor.
1. Know Your Device
Contents:
Monitor, battery set, instruction manual, setup instructions
Monitor:
1.1 Display symbols
Irregular Heartbeat Symbol ( )
When the monitor detects an irregular rhythm 
two or more times during the measurement, the 
irregular heartbeat symbol will appear on the 
display with the measurement values. 
An irregular heartbeat rhythm is defined as a 
rhythm that is 25% less or 25% more than the 
average rhythm detected while the monitor is 
measuring the systolic and diastolic blood 
pressure. 
If the irregular heartbeat symbol displays with your measurement results, 
we recommend you consult your doctor. Follow the directions of your 
doctor.
Movement Error Symbol ( )
The movement error symbol is displayed if you move your body during the 
measurement. Please remove the arm cuff, and wait 2 - 3 minutes. 
Take another measurement and remain still during measurement.
Cuff Wrap Guide Symbol (  /  )
If the cuff was wrapped too loosely, it may cause unreliable results. If the 
wrapping of cuff is too loose, the cuff wrap guide symbol “ ” appears. 
Otherwise “ ” appears. This is the function which is used as an aid in 
determining if the cuff is wrapped snugly enough. 
SYNC Symbol ( )
• When the SYNC symbol appears on the display, make sure that your 
monitor has been pairing with the smart device or transfer the data.
• When the SYNC symbol appears on the display, transfer the data before 
the measurements are deleted. If the memory is full, the monitor will be 
delete the oldest value. Maximum internal memory is for 100 
measurements.
2013 ESH/ESC Guidelines for the management of arterial hypertension
Definitions of hypertension by office and home blood pressure levels
These are from statistical values for blood pressure.
1.2 Before Taking a Measurement
To help ensure an accurate reading, follow these directions:
1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and 
eating for 30 minutes before taking a measurement. 
2. Rest for at least 5 minutes before taking the measurement.
3. Stress raises blood pressure. Avoid taking measurements during 
stressful times.
4. Measurements should be taken in a quiet place.
5. Remove tight-fitting clothing from your arm.
6. Keep a record of your blood pressure and pulse readings for your 
doctor. A single measurement does not provide an accurate indication 
of your true blood pressure. You need to take and record several 
readings over a period of time. Try to measure your blood pressure at 
the same time each day for consistency.
2. Preparation
2.1 Battery Installation
1. Remove the battery cover.
Gently slip into the gap of the 
hook and pull downward.
2. Insert 4 “AAA” batteries as 
indicated, into the battery 
compartment.
3. Replace the battery cover.
When the batteries have been securely placed in this device, the message 
and SYNC symbol below will appear on the screen:
Notes:
• When the depleted battery symbol “ ” appears on the display, turn the 
monitor off and remove all the batteries. Replace with 4 new batteries at 
the same time. Long life alkaline batteries are recommended.
• The measurement values continue to be stored in the memory even after 
the batteries are replaced.
• The supplied batteries may have a shorter life than new batteries.
• Disposal of used batteries should be carried out in accordance with the 
national/local regulations for the disposal of batteries.
2.2 Pairing Your Monitor with a Smart Device
The date and time on your monitor will automatically be set when you pair 
it with your smart device.
To begin using the “OMRON connect” app for the first time, please visit 
omronconnect.com/setup for the initial set-up instructions.
1. Turn on the Bluetooth feature of the smart device.
2. Download and install the free “OMRON connect” app on 
your smart device.
* If you already have the “OMRON connect” app, go to:
Menu > Devices > Add Device
3. Open the app on your smart device and follow the set-up 
and pairing instructions.
4. Confirm the monitor is connected successfully.
If the monitor is connected successfully to the smart device, “OK” will 
appear on the display.
If “Err” appears, refer to “Connection failure. / Data is not being 
transmitted.” in section 4.2 for more detail.
Notes: 
• Pairing this monitor with the “OMRON connect” app on your smart device 
will set the date and time on this monitor. Once you have paired with the 
“OMRON connect” app, the date and time is now setup on this monitor, 
all future measurements will include the date and time when you transfer.  
If you do not pair this monitor to set the date and time, anyࠉ
measurements you take with this monitor will not include the date and 
time.
• If you use an app other than the "OMRON connect" app, the date and 
time may not be set automatically on your blood pressure monitor.
• Please be aware that OMRON will not be responsible for the loss of data 
and/or information.
• If you use the monitor without pairing, refer to the “Using your Monitor 
without Pairing” in the “3.3 Taking a Measurement”.
3. Using the Device
3.1 Applying the Arm Cuff
Remove tight-fitting clothing from your left upper arm. 
Do not place the arm cuff over thick clothes.
1. Apply the arm cuff to your left upper 
arm.
The bottom edge of the arm cuff should be 1 
to 2 cm above the elbow. Positioning the 
display in clear view of the user makes it 
easier to read the results.
2. Securely close with the fabric 
fastener.
Notes:
• The blood pressure can differ between 
the right arm and the left arm, and the 
measured blood pressure values can be 
different. OMRON recommends to 
always use the same arm for 
measurement. If the values between both arms differ substantially, please 
check with your doctor as to which arm to use for your measurements.
• Positioning the monitor with the [START/STOP] button towards the body 
makes it easier to use the device.
3.2 How to Sit Correctly
To take a measurement, you need to be relaxed and 
comfortably seated, at a comfortable room 
temperature. 
• Sit in a chair with your legs uncrossed and your feet flat on 
the floor.
• Sit upright with your back straight.
• Sit with your back and arm being supported.
• The arm cuff should be placed on your arm at the same level as your heart.
3.3 Taking a Measurement
Notes:
• To stop the measurement, press the [START/STOP] button once to 
deflate the arm cuff.
• Remain still and do not talk while taking a measurement.
• This device will store 100 measurements in the device. Stored 
measurements cannot be viewed on the monitor's display. You will need 
to use the “OMRON connect” app to view all measurements prior to the 
most recent reading taken.
• Do not use the monitor with other electrical device simultaneously. This 
may result in incorrect operation of the monitor.
1. Press the [START/STOP] button.
The arm cuff will start to inflate.
 : The arm cuff was wrapped snugly enough.
 : The arm cuff was loose or not wrapped correctly.
Note: The cuff wrap guide result appears on the display during the 
measurement.
The following symbols indicate incorrect measurement
• If any of these symbols below are displayed, please take another 
measurement.
2. Remove the arm cuff.
3. Transfer your readings.
Please make sure your monitor is within 5 m (16 ft.) of your smart device 
and your “OMRON connect” app is active.
As soon as the measurement is complete, your reading will be 
automatically transferred to the app.
Transferring
 your readings later
1. Open the app on your smart device and then follow the instructions.
2. Press connection button to transfer your readings.
• If your readings are successfully transferred to your smart device, 
“OK” will flash on the display.
If “Err” appears, refer to “Connection failure. / Data is not being 
transmitted.” in section 4.2 for more detail.
Notes:
• When the SYNC symbol “ ” appears on the display, transfer the data 
immediately before the readings are deleted.
• When the Connection OFF symbol “ ” appears on the display, turn on 
the Bluetooth feature of your monitor. (Refer to section 3.4)
• Once you send the your readings to the app, the data cannot be resent to 
app or smart device.
4. After your readings have been transferred to the “OMRON 
connect” app, press the [START/STOP] button to turn off.
Notes:
• The monitor will automatically turn off after 2 minutes.
• Wait 2-3 minutes before taking another measurement. Waiting between 
measurements allows the arteries to return to their prior condition to 
taking the measurement.
• If you use any other app instead of “OMRON connect”, the data may not 
be displayed correctly after sending.
Data Transfer With Multiple Smart Devices
Once you transfer your data to a smart device, that data cannot be 
transferred again to the same smart device, nor to another smart device.
About Multiple Users 
This monitor is recommended for personal use. 
When this monitor is used by more than one person, the measurement data 
will be combined. 
Using your Monitor without Pairing
You can use this blood pressure monitor without pairing to a smart device.
1. Apply the arm cuff. (Refer to section 3.1)
2. Press [START/STOP] button. (Refer to step 1 to 2 in section 3.3)
3. 
Write down your SYS/DIA and Pulse readings, along with the date and time 
of your measurement on a piece of paper to keep a record of your history.
Note: When using this blood pressure monitor without a smart device, you 
will not be able to view your prior readings stored in the memory.
Upper Arm 
Blood Pressure Monitor
Model HEM-7600T
Instruction Manual
Connection symbol Battery symbol (low)
Connection OFF symbol Battery symbol (depleted)
Connection ON symbol Deflation symbol
Movement error symbol
Heartbeat symbol
(Flashes during measurement.)
SYNC symbol Cuff Wrap Guide Symbol
Irregular heartbeat symbol
Office Home
Systolic Blood Pressure ≥140 mmHg ≥135 mmHg
Diastolic Blood Pressure ≥90 mmHg ≥85 mmHg
A
BC
D
E
A. START/STOP button
B. Connection button
C. Display
D. Arm cuff
E. Battery compartment
F. Systolic blood pressure
G. Diastolic blood pressure
H. Pulse display
G
H
F
Irregular Heartbeat
Normal Heartbeat
Pulse
Blood pressure
Blood pressure
Short
Long
Pulse
 : Make sure to press the connection button to pair with the smart 
device or transfer the data before taking a measurement when 
the SYNC symbol 
“” appears.
Review the list of compatible smart devices at: 
omronconnect.com/devices
For details, visit: omronconnect.com
START INFLATING
DEFLATING
COMPLETED
DO NOT adjust medication based on measurement results from this blood 
pressure monitor. Take medication as prescribed by your doctor. Only a 
doctor is qualified to diagnose and treat high blood pressure.
Always consult your doctor. Self-diagnosis of measurement results and 
self-treatment are dangerous.
3125186-3A