EasyManua.ls Logo

Omron K3FP Series - Page 18

Omron K3FP Series
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18 Thermocouple Transducer K3FP-TS-UI
Note: K is the abbreviation for Kelvin, the unit for absolute
temperature.
Input Specifications
Note: Set the start and end temperatures using the DIP switch to
50
°C or higher.
Output Specifications
Nomenclature
Connecting
cable
Solid
wire
0.14 to 2.5 mm
2
Stranded
wire
0.2 to 2.5 mm
2
AWG 24 to 12
Wire
stripping
length
12 mm
Degree of protection IP20
Housing material PBT
Weight 54 g
Safety standards UL 508
EMC EMI:
Radiated EMI: EN 55011
EMS:
ESD immunity: EN 61000-4-2
Rated electromagnetic field immunity:
EN 61000-4-3
Burst immunity: EN 61000-4-4
Surge immunity: EN 61000-4-5
Conducted disturbance immunity:
EN 61000-4-6
Input signal
Item
Type J and K Thermocouples
(Conforms to IEC 60584-1)
Measurement range J:
150 to 1,200°C
K:
150 to 1,350°C
Minimum
measurement span
(See note.)
Min. 50
°C
Max. input signal 30 V
Output signal
Item
0 to 20 mA DC,
4 to 20 mA DC,
20 to 0 mA DC,
20 to 4 mA DC
0 to 5 V DC,
1 to 5 V DC,
0 to 10 V DC,
10 to 0 V DC
Allowable load
impedance
500
max. 10 k min.
Max. output signal 23 mA 12.5 V
Non-load voltage 12.5 V max. ---
Short-circuit current --- 10 mA
Ripple 20 mV pp max.
(500
)
20 mV pp max.
Operation during
sensor faults
0% to 105%
7
8 Supply voltage
5
6 Output signals
Translucent cover
Snap for mounting DIN rail
DIN rail bus connection
Diagnostics indicator (red)
DIP switch S3
DIP switch S2
DIP switch S1
2
3 Inputs: Type J or K thermocouple senso
r

Related product manuals