Trouble Possible cause Corrective action
The green power
indicator turns on
but the unit does not
nebulize.
The vibrating element is not
immersed in the medication.
Tilt the main unit in order to
immerse the vibrating element in the
medication.
The medication container is not
filled with medication.
Fill with medication. Refer to page
14.
The mesh is broken. Replace with a new one. Refer to
page 14.
Excessive medication is collected
on the mesh
Remove excessive medication.
Refer to page 18.
Remove visible liquid with a soft
cloth and a very light touch so as
not to damage the mesh.
The green power
indicator turns on but
it nebulizes weakly, or
is taking too long for
treatment.
Nebulization rates vary based on
the medication used.
Treatment times may vary among
medications and patients.
If the unit does not nebulize properly after checking the possible causes above, contact your
OMRON dealer.