22
Note: if the device is not cleaned and disinfected correctly and as frequently as indicated,
microorganisms may remain in the unit, causing risk of infection.
The Medication Container, Mesh Cap, Mouthpiece, Mask and Mask Adapter should be disinfected
before using the unit for the first time, after the unit has not been used for a long period of
time, or daily during normal use.
To select a method for disinfection, refer to the table below.
Parts Materials Boiling
Alcohol
Sodium
hypochlorite
Quaternary
ammonium
Chlorhexidine
Amphoteric
surfactant
Disinfecting
ethanol
Milton*
1
(0.1%)
Osvan*
1
(0.1%)
Hibitane*
1
(0.5%)
Tego*
1
(0.2%)
Mouthpiece
SEBS
||||||
Mask Adapter
PC
||||||
Medication
Container
Body: PC
Hinges: Titanium
||||||
Mesh Cap
Packing: Silicone
Mesh Holder: PC
Mesh: NiPd
||±|||
Adult Mask (PVC)
Child Mask (PVC)
Mask: PVC*
3
Band: Rubber*
4
±|*
2
|*
2
|*
2
|*
2
|*
2
Adult Mask
(SEBS) (Optional)
Child Mask
(SEBS) (Optional)
Mask: SEBS
Band: Rubber*
4
|*
2
|*
2
|*
2
|*
2
|*
2
|*
2
Vibrating
element
Titanium
±|||||
|
: applicable ± : not applicable
How to disinfect the parts