Omron MicroAIR U100 - Aerosol Performance Data; Performance Measurement Notes

Omron MicroAIR U100
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
Particle Size: *MMAD 4.5 μm
MMAD = Mass Median Aerodynamic Diameter
Medication Container Capacity: 10 ml maximum
Sound: Approx. 20 dB
Nebulization rate: **Greater than 0.25 ml/min (by weight loss)
Aerosol Output: 0.5 ml (2 ml, 1%NaF)
Aerosol Output Rate: 0.1 ml/min (2 ml, 1%NaF)
Notes:
The distribution of particle size and aerosol output are measured with 2.5% NaF aqueous solution. The
particle size distribution and aerosol output may vary by combination of product, medication and ambient
conditions such as temperature, humidity and atmospheric pressure.
Performance may vary with the mesh and medication used such as suspensions or high viscosity. See
medication supplier’s data sheet for further details.
* Measured values reflect internal data obtained by NEXT GENERATION IMPACTOR (NGI), according to
EN13544-1:2007 + A1:2009 and ISO27427:2013.
** The Nebulization rate is measured with saline 0.9% solution at 23°C at spray angle of 30° with 3 ml of
medication. It can vary with medication and ambient conditions.
0
10
20
30
40
50
60
70
80
90
100
0.1 1 10 100
Particle Size Dp (μm)
Cumulative % particle mass of sodium fluoride under size
Individual
Average
MMAD 4.5 μm
Cumulative Undersize %
Technical data