Care and maintenance
Cleaning and disinfecting
Caution: Handling the Vaporizer Head
Always wash the vaporizer head after each use.
Do not use a brush or pin to clean the vaporizer head.
When disinfecting the parts by boiling, be sure to boil them in plenty of water.
Do not boil vaporizer head together with objects other than applicable nebulizer
accessories.
Cleaning
Clean the parts after each use to remove residual medication. This will avoid inefficient
nebulization or infection.
Washable parts
Nebulizer Kit (Inhalation top, Medication tank), Mask (PVC)/(SEBS), Mouthpiece,
Nosepiece
Wash them in warm water and mild detergent (neutral detergent). Rinse them
thoroughly with clean hot tap water and allow to air dry in a clean place.
Vaporizer Head
Wash it with running water.
Non-washable parts
Compressor, Air Tube
Firstly, make sure that the power plug is unplugged from the power outlet. Wipe clean
with a soft cloth moistened with water or mild detergent (neutral detergent).
Air Filter
Do not wash or clean the air filter. If the air filter becomes wet, replace it. Damp air
filters can cause blockages.
Disinfecting
Disinfect the parts once a week. If the parts are heavily stained, replace them with new
ones.
To select a method for disinfection, refer to the table on page 14.