EasyManua.ls Logo

Omron NE-C803 - Page 19

Omron NE-C803
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
EN
Technical data for the OMRON NE-C803 Compressor with the Nebulizer kit:
Particle Size: **MMAD Approx. 3μm
MMAD = Mass Median Aerodynamic Diameter
Medication Tank Capacity: 10ml maximum
Appropriate Medication Quantities: 2ml minimum - 10ml maximum
Sound: *Less than 45dB
Nebulization Rate: *Approx. 0.30ml/min (by weight loss)
Aerosol Output: ** 0.37ml (2ml, 1%NaF)
Aerosol Output Rate: ** 0.05ml/min (2ml, 1%NaF)
Result of cascade impactor measurements for particle size**
0
10
20
30
40
50
60
70
80
90
100
0.1 1 10 100
Cumulative % particle mass of sodium fluoride under size
Particle Size Dp (μm)
Cumulative Undersize %
Mean
Individual
* Measured by OMRON HEALTHCARE Co., Ltd. Measured with saline at ambient
temperature of 23°C and humidity of 40%. It may vary with the type of medication.
** Independently measured by Prof. Dr. Hiroshi Takano, Therapeutic Systems Research
Center, Doshisha University, Kyoto, Japan, according to EN 13544-1:2007+A1:2009.
Note:
Performance may vary with medication such as suspensions or high viscosity. See
medication supplier’s data sheet for further details.

Related product manuals