7
Before Inhalation
Clean and disinfect the product’s
parts before using the product for
the first time and after the product
has not been used for an
extended period of time.
“Maintenance” ( Page 16)
“Disinfecting” (
Page 19)
Warning
When the product is being used by different people,
clean and disinfect the nebulization set, inhalation
hose, and mouthpiece after each person’s use.
Caution
Make sure the water tank contact and main unit
contact are not contaminated.
Filling the
removable water
tank with water
1
1. Grip the removable water tank’s
handle securely and lift it up to
remove it from the main unit.
Removable
water tank
handle
2. Placing your finger under the
protrusion on the medication
cup holder, remove the
nebulization set.
Protrusion
3. Fill the removable water tank
with tap water.
Fill to the optimal water level, taking care
not to exceed the maximum fill line.
Optimal
water level
Maximum
fill line
Notes:
,IWKHUHLVWRROLWWOHZDWHULQWKHUHPRYDEOHZDWHU
tank, a “
WATER LEVEL” error will occur.
Add water.
,IWKHZDWHUWHPSHUDWXUHLVWRRORZQHEXOL]DWLRQ
effectiveness may decrease.
(The recommended temperature is 26 °C.)
'RQRWSXWDQ\OLTXLGRWKHUWKDQWDSZDWHU
inside the removable water tank. It may
damage or degrade the vibrator.
5HSODFHWKHZDWHULQWKHUHPRYDEOHZDWHUWDQN
after each use. When operating the product
continuously for an extended period of time,
replace the water at least every 24 hours. See
“Using the product for extended periods of
time”. (
Page 14)