OMS ARCADIA EXT S.T.01/3R ED.0 REV.8 09/2020
20
3. COMMAND - CONTROL - WARNING
3.1. SAFETY SYSTEMS
The unit is equipped with safety devices that minimize the risk of collision during the movement of the chair, in
particular there is a safety on the basin floor water group in version, which intervenes when the tray is rotated
by the side of the chair, limiting the ascent of the chair at a defined height.
DANGER
In the version with a floor water group during the ascent of the chair check the
positioning patient on the arm to prevent the risk of crushing with the basin.
Note:
During any automatic chair movement (reset, memory placement or rinse position) by pressing
any patient chair command (from the dental unit or push off of the foot controller) the chair
stops immediately.
The movement of the chair, either at power-up that in normal use, can be prevented if one or
more safety systems are active