OMS ARCADIA EXT S.T.01/3R ED.0 REV.8 09/2020
58
IEC 61000-4-5
± 2 kV between phases and ground
± 2 kV
commercial or hospital environment
Voltage gaps, short
interruptions and voltage
variations on the input lines
dell’alimentazione
IEC 61000-4-11
<5 % U
T
(>95 % hole in U
T
per 0.5 cycle)
40 % U
T
(60 % hole in U
T
per 5 cicli)
70 % U
T
(30 % hole in U
T
per 20 cicli)
<5 % U
T
(>95 % hole in U
T
per 5S)
<5 % U
T
0.5 cycle (10mS)
40 % U
T
5 cycle (100mS)
70 % U
T
20 cycle (500mS)
<5 % U
T
5S
The quality of the mains voltage should be that of a typical
commercial or hospital environment. If the user of the
device requires continuous operation during the
interruption of the mains voltage, it is recommended to
power the device with a UPS or with batteries.
magnetic field at Network
frequency (50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
The magnetic fields at network frequency should have
characteristic levels of a typical locality in a environment
such as commercial or hospital.
NOTE: U
T
is the network tention in c.a. before the application of test level.
Manufacturer's Guide and declaration Electromagnetic immunity
The device is intended to work in the electromagnetic environment below specified. The customer or the user of the device must ensure that it is used in
this environment.
Immunity test
Level test IEC 60601
Level of
Compliance
Electromagnetic Environment-Guide
Portable and mobile RF communication equipment should not
be used near any part of the device including cables, the
recommended separation distance calculated with the equation
applicable to the frequency of the transmitter. Recommended
Distance Separation:
d = 1,17 √P
d = 1,17 √P from 80 MHz to 800 MHz