EasyManuals Logo

OneTouch Select PLUS User Manual

OneTouch Select PLUS
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #135 background imageLoading...
Page #135 background image
Detailed information
about your system
9
133
Regression Statistics
Samples were tested in duplicate on each of three test
strip lots. Results indicate that the OneTouch Select® Plus
System compares well with a laboratory method.
# of Subjects # of Tests Slope Intercept
(mmol/L)
100 600 0.97 0.12
95% CI
Slope
95% CI
Intercept
(mmol/L)
Std. Error
(S
y.x
)
(mmol/L)
R
2
0.97 to 0.98 0.05 to 0.19 0.45 0.99

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OneTouch Select PLUS and is the answer not in the manual?

OneTouch Select PLUS Specifications

General IconGeneral
Measurement Range20-600 mg/dL
Test Time5 seconds
CodingNo coding required
Battery LifeApproximately 1000 tests
Memory500 test results
Automatic Shutoff2 minutes after last action
Battery1 CR2032 lithium battery
Operating Conditions10%-90% RH
Storage Conditions10-90% relative humidity
WeightApprox. 52.5 g (with battery)

Related product manuals