OneTouch Select PLUS - User Accuracy Study

OneTouch Select PLUS
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Detailed information
about your system
9
134
Precision
Within Run Precision (300 Venous Blood Samples Tested
per Glucose Level)
Data generated using the OneTouch Select® Plus Meter.
Target
Glucose
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coecient
of Variation
(%)
2.2 2.50 0.09 3.59
3.6 3.90 0.10 2.56
6.7 7.06 0.16 2.25
11.1 11.57 0.29 2.52
19.4 19.55 0.40 2.07
Results show that the greatest variability observed
between test strips when tested with blood is 0.10 mmol/L
SD or less at glucose levels less than 5.55 mmol/L, or
2.52% CV or less at glucose levels at 5.55 mmol/L or
above.