46
Detailed information about your system
Technical specications
Reported result range 1.1–33.3 mmol/L
Calibration Plasma-equivalent
Sample Fresh capillary whole blood
Testing time 5 seconds
Assay method Glucose oxidase biosensor
Meter power source One 3.0 Volt CR 2032 lithium battery or
equivalent
Unit of measurement mmol/L
Memory 350 test results
Automatic shuto About two minutes after last action
Size 95 x 65 x 22.5 mm
Weight Approximately 60 g, with battery
Operating ranges Temperature: 10–44°C
Relative humidity: 10–90%
Altitude: up to 3048 meters
(10,000 feet)
Haematocrit: 30–55%
Battery ratings 1 x 3.0 V d.c., 60 mA
(1 x CR 2032 battery or equivalent)
direct current
Customer Care Freephone 0800 121 200 (UK) or 1800 535 676 (Ireland)