inconvenience that is improved in other software (RealTerm, Hyper Terminal
etc.).
Basic Program
1. Adjustment with the use of four trimmers. Trimmers allow to set PID
parameters, as well as the temperature setpoint.
2. The system detects a lack of connection with the thermistor and then
switches WARNING LED on. As long as the connection is lost, the driver will
not take any action and the open collector output is grounded.
3. If the measured temperature exceeds an overriding limit of +70°C in the NTC
10k versions (+120°C in the Advanced -100/+100 version), the Peltier
module is switched off.
4. If the measured temperature is out of Tmin...Tmax range, the WARNING LED
is ON and the open collector output (OC) is connected to ground. The PID
regulator works normally (unless the overriding limit mentioned before is
exceeded). In the Basic version, Tmin and Tmax are 0°C and 70°C