11
Selection of Division
• To change the max capacity, in the settings mode press [MODE] until
the display reads “D0.00##” ex. “d0.0002” means the division is 0.2g
• The default number is the actual division of the scale
• Use the [UNIT] key to toggle between options
• Finally, press [MODE] key to conrm it, and the display will show the
next menu
CALIBRATION
When to calibrate
• When the scale is initially installed
• When the scale’s location has changed
• After the scale has not been used for a long period of time
• If ambient temperatures have changed greatly
How to calibrate
• Enter the settings mode by pressing [MODE] and [TARE], while you turn
on the power
• Press the [MODE] key 6 times until the display reads “E#.0000”
• Press [UNIT] to choose the calibration weight (we suggest to use 2/3 of
the weight of the scales capacity)
ex. “E5.0000” would mean you calibrate with a 5kg weight
• You can change the weight by using the [ZERO] key to go up and the
[TARE] key to go down and the [UNIT] key to change to the next
sequential number
• Conrm your weight by pressing [MODE] key
• It will display ZeroAD value (range from 10000-60000)
• Press [MODE] key again and it will display “-Load-” for two seconds,
please place your weight that is the same as the one you just entered
on the scale within two seconds!
• Press [MODE] to conrm