11
Selection of Division
To change the max capacity, in the settings mode press [MODE] until
the display reads “D0.00##” ex. “d0.0002” means the division is 0.2g
The default number is the actual division of the scale
Use the [UNIT] key to toggle between options
Finally, press [MODE] key to conrm it, and the display will show the
next menu
CALIBRATION
When to calibrate
When the scale is initially installed
When the scale’s location has changed
After the scale has not been used for a long period of time
If ambient temperatures have changed greatly
How to calibrate
Enter the settings mode by pressing [MODE] and [TARE], while you turn
on the power
Press the [MODE] key 6 times until the display reads “E#.0000”
Press [UNIT] to choose the calibration weight (we suggest to use 2/3 of
the weight of the scales capacity)
ex. “E5.0000” would mean you calibrate with a 5kg weight
You can change the weight by using the [ZERO] key to go up and the
[TARE] key to go down and the [UNIT] key to change to the next
sequential number
Conrm your weight by pressing [MODE] key
It will display ZeroAD value (range from 10000-60000)
Press [MODE] key again and it will display “-Load-” for two seconds,
please place your weight that is the same as the one you just entered
on the scale within two seconds!
Press [MODE] to conrm