What to do if my Optimanufacturing T60-1001-C8-1A Measuring Instruments tensiometer reads low?
- DdsantiagoAug 12, 2025
If your Optimanufacturing Measuring Instruments tensiometer reads lower than expected, the issue could stem from several causes. It might be due to a weak trigger spring (29), in which case you should replace the trigger spring and calibrate the device. Another potential cause is dirt or foreign matter lodged between the frame (41) and sectors (24); cleaning and reassembling the parts may resolve this. Worn-out sectors (24) or riser block (5) could also be the culprit, requiring replacement of these components followed by calibration. Finally, a faulty dial indicator (35) could be the reason, which means you'll need to replace the dial indicator and calibrate.