SOP-TN-002 Rev. Q
T60 Operations and Service Manual
OPERATION & SERVICE MANUAL FOR
CABLE TENSIOMETER
T60 SERIES
PART NUMBERS:
T60-1001-C8-1A (NSN: 6635-00-530-1128)
T60-1001-C9-1A (NSN: 6635-00-530-1129)
T60-1002-C8-00 (BRITISH MODEL FOR CWT CABLES)
T60-1002-C9-00 (BRITISH MODEL FOR CWT CABLES)

Questions and Answers

  • T
    Thomas AndrewsAug 12, 2025
    What to do if my Optimanufacturing T60-1001-C8-1A Measuring Instruments tensiometer reads low?
    • D
      dsantiagoAug 12, 2025
      If your Optimanufacturing Measuring Instruments tensiometer reads lower than expected, the issue could stem from several causes. It might be due to a weak trigger spring (29), in which case you should replace the trigger spring and calibrate the device. Another potential cause is dirt or foreign matter lodged between the frame (41) and sectors (24); cleaning and reassembling the parts may resolve this. Worn-out sectors (24) or riser block (5) could also be the culprit, requiring replacement of these components followed by calibration. Finally, a faulty dial indicator (35) could be the reason, which means you'll need to replace the dial indicator and calibrate.
  • K
    Kathy MorganAug 20, 2025
    What to do if the handle latch on my Optimanufacturing T60-1001-C8-1A Measuring Instruments won't lock?
    • R
      Richard HopkinsAug 20, 2025
      If the handle latch on your Optimanufacturing Measuring Instruments isn't locking the handle in the closed position, it's likely that the handle latch spring (39) is either weak or broken. Replace the latch spring to fix this. Calibration is not required after this replacement.
  • J
    Jennifer GonzalesSep 1, 2025
    Why won't the pointer lock function on my Optimanufacturing Measuring Instruments?
    • J
      jmaySep 1, 2025
      If the pointer lock on your Optimanufacturing Measuring Instruments is not working, the switch spring (30) is likely broken. Replace the switch spring and then calibrate the instrument to resolve this issue.