Ormed ATRTOMOT-K1 - 3.2 Mechanics; 3.3 Others

Ormed ATRTOMOT-K1
17 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
Checksum
error
chipcard:
-
>
Replace
the
chipcard
(Pos.
21)
38 Unknow
chipcard
error:
-
>
Replace
the
hand
held
programming
unit
(Pos.
20)
39 A
chipcard
of
another
ARTROMOT®
product
got
used:
-
>
Format
the
chipcard;
Press
START
-> Replace the chipcard (Pos. 21)
40 Unkown
chipcard:
-
>
Replace
the
chipcard
(Pos.
21)
41 Chipcard
is
missing:
No chipcard is insert
-
>
Insert
a
chipcard
-> Replace the chipcard (Pos. 21)
42 Chipcard
write
error:
The chipcard is not writable
-
>
Replace
the
chipcard
(Pos.
21)
-> Replace the hand held programming unit (Pos. 20)
43 Chipcard
read
error
Chipcard is not readable.
-
>
Replace
the
chipcard
(Pos.
21)
-> Replace the hand held programming unit (Pos. 20)
44 Chipcard
verify
:
error
-
>
Replace
the
chipcard
(Pos.
21)
-> Replace the hand held programming unit (Pos. 20)
45 Wrong
product
combination:
Mixup between non compatible device and hand held
programming unit
-
>
Use
the
correct
hand
held
rogramming
unit
(Pos.
20)
46 Handset
error
internal
communication:
Invalid interchip communication inside the hand held
programming unit
-
>
Replace
the
hand
held
programming
unit
(Pos.
20)
47 Internal
communication
error
motor
control:
Internal communication error motor control.
-
>
Repl
ace
motor
control
(Pos.
25)
48 User
stoped
the
special
function
4
49
Unknown
error
in
the
motor
control:
-
>
Repl
ace
motor
control
(Pos.
25).
Chipcard is missing
-> Insert a chipcard
-> Replace the chipcard (Pos. 21)
Chipcard error
-> Replace the chipcard (Pos. 21)
->
Replace the hand held programming unit (Pos. 20)
3.2
Mechanics
The
movable
screws
should
not
be
completely
unscre
-
wed
when adjustments are being made. Make sure
that the movable screws are tightened for operation
and transport.
The frame is unstable:
Possible cause: Bolt / screws missing or loose. Tighten
the screws / bolts.
3.3
Others
Do
not
clean
the
housing
or
the
support
with
grease
or
oil.
No solvents may be used when cleaning the
ARTROMOT®-K1.