User’s Manual (Confidential)
OsteoSys Co., Ltd.
Model Designation : SONOST 3000 DOC. No. : OT06-2R7124
27
Phantom’s setting
Apply the ultrasound transmission to the phantom to the area which will be in
contact with both probes when the phantom is on the foot rest. The amount of
gel should be enough to cover the area of the phantom that comes in contact
with the probes.
Put the phantom on the foot supporter 1 (don’t use foot supporter 2 for Daily
Test) in the measuring position of device.