What to do if test results seem incorrect on otometrics Measuring Instruments?
- IisaiahdavisAug 1, 2025
If your test results seem incorrect, follow these steps: 1. Open AURICAL HIT and remove the hearing instrument and the coupler microphone. 2. Point the reference microphone towards the area of the coupler microphone socket in AURICAL HIT. 3. Close the lid. 4. Run the test. 5. Look at the curve at 90 dB SLP. It must be as flat as possible. 6. If it is not, make a Validation Report, recalibrate, and repeat the sequence above. 7. If test results still seem incorrect, replace the relevant microphone and calibrate the new microphone.