1.Safety Information
6. Technical Specifications
Function Range
[2]
Resolution Accuracy: ± (% of reading + LSB)
DC Voltage
50.000 mV 0.001 mV 0.1% + 10
500.00 mV 0.01 mV 0.05% + 5
5.0000 V 0.0001 V 0.05% + 5
50.000 V 0.001 V 0.05% + 5
500.00 V 0.01 V 0.1% + 5
1000.0 V
[3]
0.1 V 0.1% + 10
True RMS
AC Voltage
[4]
500 mV – 750 V
20 Hz – 45 Hz 1% + 30
45 Hz – 65 Hz 0.5% + 30
65 Hz – 1 kHz 0.7% + 30
DC Current
500 uA 0.01 uA 0.15% + 20
5000 uA 0.1 uA 0.15% + 10
50 mA 0.001 mA 0.15% + 20
500 mA 0.01 mA 0.15% + 10
5 A 0.0001 A 0.5% + 10
10 A
[5]
0.001 A 0.5% + 10
True RMS
AC Current
[6]
500 uA – 500 mA
/
0.5% + 20
5 A – 10 A 1.5% + 20
Resistance
[7]
500 Ω 0.01 Ω 0.15% + 10
5 kΩ 0.0001 kΩ 0.15% + 5
50 kΩ 0.001 kΩ 0.15% + 5
500 kΩ 0.01 kΩ 0.15% + 5
5 MΩ 0.0001 MΩ 0.3% + 5
50 MΩ 0.001 MΩ 1% + 10
Diode 3.0000 V 0.0001 V 1% + 10
Continuity 1000 Ω 0.1 Ω Adjustable threshold
Frequency 10.000 Hz – 60 MHz
[8]
/ ± (0.2% + 10)
Capacitance
[9]
50 nF – 500 uF / 2.5% + 10
5 mF – 50 mF / 5% + 10
Temperature K type, PT100
Display 55,000
Record Interval 15 mS – 9999.999 S
Record Length 1,000 points
[1]
Specifications are for 30-minute warm-up, "Low" measurement rate and calibration temperature
18
℃
-G28
℃
.
[2] 10% over range on all ranges, except 1,000 V DCV, 750 ACV, 10 A DCI, 10 A ACI, 50 MΩ resistance, and
50 mF capacitance.
[3] For each additional volt over ± 500 VDC add 0.02 mV of error.
[4] Specifications are for amplitude of sine wave input > 5% of range. 0.1% errors will be added when the
range of input sine wave is 1% to 5%.
[5] 30 seconds OFF after 30 seconds ON is recommend for the continuous current that higher than DC 7 A or
AC RMS 7 A.
[6] Specifications are for amplitude of sine wave input > 0.5% of range.
For inputs from 1% to 5% of range,
add 0.1% of range extra error.
[7] Specifications are for 2–wire ohms using the relative operation of math. Without relative operation, add
±0.20 Ω additional error in 2-wire ohms function.