19
7. Cleaning, Care and Maintenance
Cannula Replacement
Your nasal cannula should be replaced on a regular basis. Consult with your physician
and/or equipment provider and/or cannula manufacturer’s instructions for replacement
information. A single lumen cannula up to 25 feet in length is recommended to ensure proper
breath detection and oxygen delivery.
Case Cleaning
You may clean the outside case using a cloth dampened with a mild liquid detergent
(such as Dawn
TM
) and water.
Filter Cleaning and Replacement
The particle lters must be cleaned weekly to ensure the ease of air ow. Remove lters from the
front and back of the device. Clean the particle lters with a mild liquid detergent
(such as Dawn
TM
) and water; rinse in water and dry before reuse.
To purchase additional particle lters contact your equipment provider.
Output Filter
The output lter is intended to protect the user from inhalation of small particles in the product
gas ow. The OxyGo NEXT includes an output lter conveniently located behind the
removable cannula nozzle tting.
Under normal conditions the output lter could last the life of the product.
DC Input Cable Fuse Replacement
The Cigarette Lighter DC power plug contains a fuse. If the DC input cable is being used with
a known good power source and the unit is not receiving power, the fuse may need to be
replaced.
To replace the fuse, follow these instructions and refer to the photograph below.
Remove the tip by unscrewing the retainer. Use a tool if necessary.
Remove the retainer, tip and fuse.
The spring should remain inside the Cigarette Lighter Adapter housing. If the spring is
removed, make sure to replace the spring rst before inserting the replacement fuse.
Install a replacement fuse, BUSS MDA -12 (1400-1051) and reassemble the tip.
Ensure the retainer ring is properly seated and tightened.
Standard and Optional Accessories
OxyGo NEXT single battery 1400-3010-8
OxyGo NEXT double battery 1400-3010-16
Carry Bag 1170-3410
Backpack 1170-3420
External Battery Charger 1400-3030
AC Power Supply 1400-3040
DC Power Cable 1400-1050