EasyManua.ls Logo

OxyGo Next - Specifications

OxyGo Next
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
8. Specications
Dimensions:
w/ 8-cell battery
w/ 16-cell battery
L / W / H: 7.19 in. / 3.26 in. / 7.11 in.
L / W / H: 7.19 in. / 3.26 in. / 8.15 in.
L / W / H: 7.19 in. / 3.26 in. / 9.03 in.
Weight: 4.77 pounds (includes single battery)
Noise: 38 dBA at setting 2
Maximum Sound Power of 60 dBA and Maximum
Sound Pressure level of 50 dBA per ISO 80601-2-69
Warm up time: 2 minutes
Oxygen Concentration**: 90% - 3% /+ 6% at all settings
Flow Control Settings: 6 settings: 1 to 6
Maximum Outlet Pressure < 28.9 PSI
Power: AC Power Supply:
DC Power Cable:
Rechargeable Battery:
AC Input: 100 to 240 VAC
50 to 60 Hz
Auto-Sensing: 2.0-1.0A
DC Input: 13.5-15.0VDC,10A Max.
Voltage: 12.0 to 16.8 VDC (±0.5V)
Battery Duration*: Up to 6.5 hours with single battery
Up to 13 hours with double battery
Battery Charging Time: Up to 3 hours for a single battery
Up to 6 hours for a double battery
Environmental Ranges Intended for
Use:
Temperature: 41 to 104˚F (5 to 40˚C)
Humidity: 0% to 95%, non-condensing
Altitude: 0 to 10,000 ft (0 to 3048 meters)
Environmental Ranges Intended for
Shipping and Storage:
Temperature: -13 to 158˚F (-25 to 70˚C)
Humidity: 0% to 95%, non-condensing
Store in a dry environment
Altitude: 0 to 10,000 ft (0 to 3048 meters)
Transportation: Keep Dry, Handle With Care
*Battery time varies with ow setting and environmental conditions
** Based on atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C)
Contains Transmitter Module IC: 2417C-BX31A. Contains FCC ID: N7NBX31A
This device complies with Part 15 of the FCC rules. Operation is subject to the following two
conditions: (1) this device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause undesired operation.
Classication:
IEC Class II Equipment
Type BF Applied Part
IP22 Drip Proof
Not suitable for use in the presence of a
ammable anesthetic mixture
with air or with oxygen or nitrous oxide.
Continuous Operation

Table of Contents