6.4 EMC compliance table
19
FR
VIII. TABLEAU DE CONFORMITE CEM
EMC Compliance according to IEC 60601-1-2 (2007)
The NOCOSPRAY 2 is intended for use in the electromagnetic environment specified below.
The customer or the user of the NOCOSPRAY 2 should assure that it is used in such an environment.
Emission test Standard Compliance
Electromagnetic environment
Emissions RF CISPR 11 Group 1 The NOCOSPRAY 2 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Emissions RF CISPR 11 Class B The NOCOSPRAY 2 is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low
voltage power supply network that supplies buildings used for domestic
purposes.
Harmonics emissions IEC 61000-3-2 N/A
Voltage/Flicker test IEC 61000-3-3 N/A
The NOCOSPRAY 2 is intended for use in the electromagnetic environment specified below. The customer or the user of the NOCOSPRAY 2
should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance level
Electromagnetic environment
Electrostatic discharges
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30 %.
Electrical Fast Transients
IEC 61000-4-4
2 kV for
power
supply lines
1 kV for
input/output
lines
2 kV for power
supply lines
Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV line(s) to
line(s)
2 kV line(s) to e
1 kV line(s) to
line(s)
2 kV line(s) to ear
Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 % U
T
(60 % dip in UT)
for 5 cycles
70 % U
T
(30 % dip in UT)
for 25 cycles
<5 % U
T
(>95 % dip in
U
T)
for 5 s
<5 % U
T
(>95 % dip in UT)
for 0,5 cycle
40 % U
T
(60 % dip in UT)
for 5 cycles
70 % U
T
(30 % dip in UT)
for 25 cycles
<5 % U
T
(>95 % dip in UT)
for 5 s
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the system requires continued operation
during power mains interruptions, it is recommended that the system be
powered from an uninterruptible power supply or a battery.
U
T is the a.c. mains voltage prior to application of the test level.
Magnetic field at power
frequency
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.