2.4 Material information
The individual product components are made from the follow-
ing materials:
Product component Material
Inspiratory valve Polypropylene, silicone
Nozzle insert Polypropylene
Nebuliser lower part Polypropylene
Mouthpiece (with
exhalation valve)
Polypropylene, thermoplastic elastomer
Connection tubing Polyvinyl chloride
Tubing endpiece Thermoplastic elastomer
PARIchildmasksoft Polypropylene, thermoplastic elastomer
Elastic band Synthetic rubber
2.5 Maintenance
The compressor is maintenance free.
2.6 Operating life
The individual product components have the following expec-
ted lifetimes:
Product component Operating life
Compressor Approx. 500operating hours (this
corresponds to max. 3years)
If the compressor is still in use
after this time, have the
compressor tested. To do this,
contact the manufacturer or
distributor.
Nebuliser (all components
except the connection tubing)
300disinfections, max. 1year
Connection tubing max. 1year
PARIchildmasksoft 300disinfections, max. 1year
– 16 –