If applicable, detach the
mouthpiece from the nebuliser.
Attach the mask to the
nebuliser.
Filling the nebuliser
Insert the nebuliser in the holder on the compressor in-
tended for this purpose.
If applicable, detach the inspiratory valve from the nebuliser.
Pour the required quantity of
inhalation solution into the top
of the nebuliser.
Note the minimum and
maximum fill volumes
[see:General nebuliser data,
page36]. If the nebuliser
contains too little or too much
liquid, the nebulisation and
consequently the therapy will
be less effective.
Fit the inspiratory valve onto
the nebuliser.