4 REPROCESSING
The product components must be cleaned thoroughly immedi-
ately after each use and disinfected at least once a day.
If the nebuliser is used in professional environments, follow
the information on reprocessing included in the appendix at the
end of these instructions for use.
4.1 Preparation
Detach the tube from the nebuliser.
Detach the mouthpiece from the nebuliser.
Make sure that all residual volume is removed from the
nebuliser.
Dismantle the nebuliser into its individual parts.
Carefully pull the blue exhala-
tion valve out of the slot in the
mouthpiece. The exhalation
valve must still be attached to
the mouthpiece.
4.2 Cleaning
The connection tube cannot be cleaned or disinfected. Re-
garding care of the connection tube, follow the instructions
in the corresponding section [see:Care of the connection
tube, page22].
Briefly rinse all parts used in running drinking water
beforehand.
Place all disassembled com-
ponents in warm drinking water
with a little dishwashing liquid
for about 5min.
Rinse off all parts thoroughly in drinking water.
Shake the water out of all of the parts.
– 20 –