4.7 Storage
Store this product as described below:
• Wrap all individual components in a clean, lint-free cloth
(e.g., a tea towel).
• Store this product in a dry, dust-free place.
5 TROUBLESHOOTING
Fault Possible cause Remedy
No aerosol is
coming out of
the nebuliser.
The nebuliser
nozzle insert is
blocked.
Clean the nebuliser.
The connection
tubing is not con-
nected properly.
Check that the tubing connect-
ors are connected firmly to the
compressor and the nebuliser.
The connection
tubing is leaking.
Replace the connection tubing.
6 TECHNICAL DATA
6.1 General nebuliser data
Size
5
10cm×9.5cm×4.5cm
Weight
5
27.5g to 29.5g
Operating gases Air
Minimum compressor flow 3.0l/min.
Minimum operating pressure 0.5bar / 50kPa
Maximum compressor flow 6.0l/min.
Maximum operating pressure 2.0bar / 200kPa
Minimum fill volume 2ml
Maximum fill volume 8ml; PARILCSTAR: 6ml
5) Without mouthpiece; unfilled.
– 23 –