Mechanical
cleaning with
disinfection:
Alkaline cleaning agent, e.g., neodisher
®
MediClean forte (Dr.Weigert) in conjunction with
neutralising agent, e.g., neodisher
®
Z
(Dr.Weigert)
Equipment:
– Cleaning device and disinfector in conform-
ance with DINENISO15883, e.g.,
RDGG7836CD (Miele)
– Special baskets for Miele instrument dish-
washer
– Compressed air source for blowing dry
Vario TD programme or comparable valid
programmes
3. Steam sterilisation
not usable
4. Visual inspection & storage
Inspect all individual parts. Replace any broken, misshapen or
seriously discoloured parts.
Storage location:
– dry
– dust-free
– protected from sources of contamination
optional: Use sterile packaging
Processing limits
max. 50processing cycles
– 30 –