Parker NITROSource N2-45P - User Manual

Parker NITROSource N2-45P
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Nitrogen Gas Generator
NITROSource N2-20P - N2-80P
Original Language
User Guide
aerospace
climate control
electromechanical
filtration
fluid & gas handling
hydraulics
pneumatics
process control
sealing & shielding
EN

Questions and Answers

  • W
    wendy02Aug 7, 2025
    Why is there no or low gas outlet pressure in my Parker Laboratory Equipment?
    • L
      Luis FoxAug 7, 2025
      There could be several reasons for low or no gas outlet pressure. The generator might be overdue for service, so service the generator. There could be an internal or external gas leak, so check and rectify. Also, the inlet pressure might be too low; ensure the pressure meets the required specification.
  • K
    Katherine HuberAug 12, 2025
    What to do if the LCD and status indicators are not illuminated on my Parker NITROSource N2-45P even when power is connected?
    • M
      martinezsarahAug 13, 2025
      If the LCD and status indicators are not illuminated, even with power connected, the fuse may have blown, so replace the fuse. Alternatively, the ribbon cable might be disconnected, so reconnect the ribbon cable. Another possibility is that the power is disconnected, so reconnect the power.
  • C
    Cody AndrewsAug 17, 2025
    How to troubleshoot low inlet pressure in Parker NITROSource N2-45P Laboratory Equipment?
    • N
      Neil DawsonAug 18, 2025
      If you are experiencing low inlet pressure, the compressor or ring main pressure may be low, so check and rectify. The inlet valve may not be open, so check and rectify. There could also be a defect on the pre-treatment package; in this case, refer to the pre-treatment manual.
  • S
    Steven PeckAug 22, 2025
    What to do if there is a high oxygen concentration in my Parker NITROSource N2-45P?
    • A
      Amy BurkeAug 22, 2025
      A high oxygen concentration may be due to a defective oxygen cell, so replace it. Another cause could be a leak in the system piping, so check and rectify.
  • A
    Anthony SuarezAug 27, 2025
    How to reduce excessive noise or vibration from Parker Laboratory Equipment?
    • J
      Jordan SmithAug 27, 2025
      Excessive noise or vibration could be due to a silencer that is loose or defective, so check and rectify. Alternatively, the solenoid valve may be worn or the coil loose, so check and replace if required.
  • V
    Victor RobertsSep 1, 2025
    What causes high outlet pressure in Parker Laboratory Equipment?
    • A
      Adam CollierSep 1, 2025
      High outlet pressure can be caused by a defective outlet regulator, so reset or replace it.

Summary

Safety Information

Markings and Symbols

Explains safety markings and international symbols used on the equipment and in the manual.

Description

Technical Specification

Provides detailed performance data, including flow rates and purity levels for different models.

Installation and Commissioning

Compressed Air Pre-treatment

Details requirements and recommendations for air pre-treatment systems to ensure optimal generator operation.

Mechanical Installation

Covers general requirements for pipe-work, material selection, and routing for safe and effective installation.

Making the Connections

Guides the user through connecting the various ports, regulators, and pipe-work to the generator.

Electrical Installation

Outlines the procedures and requirements for safely connecting the generator to the electrical supply.

Customer Connections

Details how to connect external signals, alarms, and supplies like mains voltage and dryer to the generator.

Operating the Generator

Starting the Generator

Step-by-step instructions for safely starting the nitrogen generator and initiating the start-up sequence.

Stopping the Generator

Details the correct procedure for shutting down the generator and returning it to standby mode.

Oxygen Sensor Calibration

Provides a detailed procedure for calibrating the oxygen sensors to ensure accurate purity readings.

Preventative Maintenance

Troubleshooting

Fault Identification and Remedies

Identifies common faults, their probable causes, and the recommended remedies for resolving issues.

Parker NITROSource N2-45P Specifications

General IconGeneral
BrandParker
ModelNITROSource N2-45P
CategoryLaboratory Equipment
LanguageEnglish