Handy Polaris – Instruction Manual
Screw the ring in place. Tighten fairly tightly, but the membrane must not wrinkle. If this
happens try again with a new membrane and O-ring. An O-ring can only be used once.
Fill the cap with type 1 electrolyte and screw it slowly onto the probe. Excess electrolyte will
dribble from the thread. Calibrate the probe. Re-calibrate after a few hours, since the probe
will take a little time to settle down after renovation.
Page 8 of 12 220416IM-Iss03 Issue Date 22/04/2015