- 14 -
<Pre-usage>
User preparation
No. Content
1
The device is designed and validated for measurements on arm diameter of 19 to 32 cm. Arm
diameter higher or lower than the range may lead to incorrect measurements.
2 Individuals should not eat, drink, smoke, exercise or take a bath/shower in 30 min before measuring.
3 Measurement should be done with emptied bladder.
4
Individuals should take off thick clothes, such as coat and sweater, etc. Thin clothes like
shirts not disturbing the measurement. Expose the left upper arm for measurement. Do not
compress the arm when rolling up the sleeve.
5
Individuals should rest statically for more than 10 min, and perform the measuring in a quiet
and appropriate (Temperature 5 to 40˚C, relative humidity not more than 80%) environment.
6 Please maintain both the body and mood in a peaceful state when performing the measuring.
Check List
Please check the following before performing the measuring.
Before
turning
on power
supply
No. Contents
1 Ensure there is no damage or deformity, etc. on the device and its accessories.
2
Ensure there is no moisture condensation on the device. If so, please dry it
thoroughly first.
3 Ensure the device is placed steadily with sucient space around.
4 Ensure there are no objects in the arm cylinder.
5
Please connect the device to power supply as instructed in this manual, with the
designated AC adapter or AA baeries.
6 Ensure the device is correctly connected to the numerical keypad and printer.
After
turning
on power
supply
1 Conrm there is no abnormal sound from the device.
2 Ensure there is no smoke or unusual odor coming out from the device.
3 Ensure there is no excessive heat in the device.
4 Ensure the display of LCD screen on the device and numerical keypad is normal.
5 Ensure the buttons on the device, numerical keypad and printer are normal.
6
Ensure the emergency release button is in the pop-up status. For details see
"handling of emergency situations in page 34.
7
Ensure there is normal compression and air release in the cuff when pressing the
button.