443B101/443B102 12
Manual 19854 Rev. D
ECN 50523
Command: OUTS
Purpose: Sets the output sensitivity of the amplifier.
Response: The module responds with a “0” to tell the calling process the command has
been received.
Command String: XYCOZOUTSXX.XX
Description: X— Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y— Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z— Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying “C02” addresses a 443B102. (Range: 3 ASCII printable
characters)
OUTS— Commands the module to set the output sensitivity
XX.XX— 5 character output sensitivity. (Range: ASCII printable characters, 0
through 9)
Response String: “0”
Description: This return indicates that the command was received by the card.
Example: Command String: “02C02OUTS1.001”
Response String: “0”
The output sensitivity of the 443B102 in rack 0, and slot 2 will be set to 1.001.