443B101/443B102 17
Manual 19854 Rev. D
ECN 50523
Command: LOWF
Purpose: Sets the low frequency response of the amplifier.
Response: The module responds with a “0” to tell the calling process the command has
been received.
Command String: XYC0ZLOWFZ
Description: XRack number of the target module (Range: ASCII printable characters, 0
through 3)
Y Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying “C02” addresses a 443B102.
LOWF Commands module to change the low frequency response.
Z specifies the low frequency response as follows:
“1” sets the low frequency response to 0.2 Hz.
“2” sets the low frequency response to 2.0 Hz
“3” engages the medium time constant.
“4” engages the long time constant.
Response String: “0”
Description: This return indicates that the command was received by the card.
Example: Command String: “06C02LOWF3”
Response String: “0”
The 443B102 in rack 0 and slot 6 is set to medium time constant mode.