443B101/443B102 29
Manual 19854 Rev. D
ECN 50523
using English units in integration mode.
XX.XX pC/m/s^2 Transducer sensitivity for a charge sensor
using SI units in integration mode.
Low Frequency
Response 0.2 Hz Low frequency response 0.2 Hz.
2.0 Hz Low frequency response 2.0 Hz.
Med TC Medium time constant .
Long TC Long time constant.
S Int 1 Hz Single integration to 1 Hz
S Int 10 Hz Single integration to 10 Hz
D Int 1 Hz Double integration to 1 Hz
D Int 10 Hz Double integration to 10 Hz
Low Pass
Filter 0.1 kHz Low pass filter set to 100 Hz.
1.0 kHz Low pass filter set to 1.0 kHz.
3.0 kHz Low pass filter set to 3.0 kHz.
10 kHz Low pass filter set to 10 kHz.
30 kHz Low pass filter set to 30 kHz.
100 kHz Low pass filter set to 100 kHz.
Integration
Units SI SI units will be used for input and output
Eng English units will be used for input and output sensitivities in
integration mode.
Reference Ref On The internal sinusoidal reference is on.
Ref Off The internal sinusoidal reference is off.
Overvoltage OV=1 The 443B is overloaded.
OV=0 The 443B is not overloaded.
Fault
(only sent in
ICP mode) Fault=1 There is an open or short at the input.
Fault=0 The sensor bias is within the acceptable range.
Zero Lock
(only sent in
long TC charge
mode) Zero Lock On The reset is clamped on.
Nothing is sent if zero lock is off.
Example: Command String: “02C02STAT
Response String: “ICP 2mA;10.00 mV/unit;
1.023 mV/unit;2.0 Hz;10kHz; SI;