443B101/443B102 32
Manual 19854 Rev. D
ECN 50523
Command: MTED
Purpose: Returns more data stored in the TEDS sensor .
Note: Sensors programmed with the “Accelerometer, transfer function v0.91”
template (UTID 116225) contain more data than can be returned by the TEDR
command so MTED must be used also to retrieve all of the stored data.
Response: The module responds with a string of data delimited with semicolons.
Command String: XYCOZMTED
Description: X Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying “C02” addresses a 443B102.
MTED Commands the module to return more data from the TEDS sensor.
Response String: Low pass cut-off frequency; Resonance Frequency; Quality factor; Amplitude
Slope; Temperature coefficient; Reference temperature;
Example: Command String: “04C01MTED”
Return String: F lp 10011; Fres 100336; Mounted Q 10.8; Amp
Slope 1.002; Temp Coeff 0.236; Ref Temp 25.0;
The 443B101 in rack 0 and slot 4 is commanded to return more data stored in
the TEDS sensor connected to its input.