443B101/443B102 33
Manual 19854 Rev. D
ECN 50523
Command: TEDD
Purpose: Returns raw hex data stored in the TEDS device.
Response: The module responds with the CRC followed by a 32 byte string of hex data.
Command String: XYCOZTEDD
Description: X— Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y— Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z— Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying “C02” addresses a 443B102.
TEDD— Commands the module to return raw hex data stored in the
TEDS device.
Response String: CRC followed by 32 bytes of hex data.
Example: Command String: “04C01TEDD”
Return String: C917D014D00E942200005C12EC64352D
8701000000000000000000000000000000
(where C9 is the CRC byte)
The 443B101 in rack 0 and slot 4 is commanded to turn return the raw hex data
stored in the TEDS device.