443B101/443B102 35
Manual 19854 Rev. D
ECN 50523
Command: TED0
Purpose: Disables the auto TEDS normalization feature.
Response: The module responds with a “0” to tell the calling process the command has
been received.
Command String: XYC0ZTED0
Description: XRack number of the target module (Range: ASCII printable characters, 0
through 3)
Y Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying “C02” addresses a 443B102.
TED0 disables the auto TEDS normalization feature.
Response String: “0”
Description: This return indicates that the command was received by the card.
Example: Command String: “06C02TED0”
Response String: “0”
The 443B102 in rack 0 and slot 6 is commanded to disable the auto TEDS
normalization feature.