443B101/443B102 36
Manual 19854 Rev. D
ECN 50523
Command: TEDU
Purpose: Writes to the user data field of a TEDS sensor (13 characters maximum).
Response: The module responds with a “0” to tell the calling process the command has
been received.
Command String: XYC0ZTEDU
Description: X—Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y— Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z— Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying “C02” addresses a 443B102.
TEDU— tells the module to write to the user data field of a TEDS sensor.
Response String: “0”
Description: This return indicates that the command was received by the card.
Example: Command String: “06C02TEDU13 Characters”
Response String: “0”
The 443B102 in rack 0 and slot 6 is commanded to write The string “13
Characters” to the user data field of a TEDS sensor.