EFP-500 M0001-006 07/26/2011 - 9 -
Connection for
Aux-Water or
Elevator Wire
Channels
4 Operation
The EFP-500 is designed to flush the internal channels of Olympus flexible endoscopes in
accordance with Olympus reprocessing instructions.
NOTE: The EFP-500 is intended for use with the endoscope manufacturer’s
manual cleaning instructions. The EFP-500 is a substitute for
cleaning activities that require the use of a manual syringe prior to
manual high-level disinfection or the endoscope being placed in an
automated endoscope reprocessor (AER) for disinfection.
Start-Up
The unit may be placed on a countertop. However, it is preferred that the unit be mounted on the wall
to prevent inadvertent movement of the device. Plug the power cord into the right side of the EFP-
500 and into a 115V GFCI protected wall receptacle.
WARNING
Since the EFP-500 is used around water, it must be
plugged into a GFCI (Ground Fault Circuit Interrupter)
outlet.
Inlet Hose
Connection
Connection for
Suction/Instrument &
Air/Water Channels
Air/Water Channel
Connection
Auxiliary Water Channel
Connection
Suction/Instrument
Channel Connection
Container
CLEN218